Atlas
- Amélioration du comportement pléthysmographique (forme d’onde SpO₂) pendant la RCP afin que les pulsations restent visibles pendant les compressions et puissent refléter à la fois l’activité cardiaque intrinsèque et les pulsations générées par la RCP.
- Introduced a dedicated Pulse Rate (PR) value separate from Heart Rate (HR), improving handling of pulseless states, CPR, low perfusion, and probe disconnection while keeping PR aligned with HR by default
Camera
Exported videos now include ECG, SpO₂, and ETCO₂ waveforms as shown in live preview
Corpuls
- Improved AED analysis: removed incorrect “Patient Movement Detected” message and enhanced shock decision timing using ECG analysis blocks
- Added CPR depth support with manikin-specific ranges
- Improved CO₂ and apnea alarm behavior, including correct triggering and translations
- Added SpO₂ probe-off alarm with localized messaging
- Fixed metronome audio and defibrillator behavior, including QRS marker display and power-on logic
- Ensured alarm notifications clear correctly after Autolimits adjustments
CPR Band / CPR Puck
- Improved Bluetooth connectivity and compatibility with newer devices
CPR Console
- Fixed compression depth unit conversion and display accuracy
Facilitator/Control
- Improved ECG waveform behavior: heart rate now correctly updates or persists depending on waveform type
- Corrected passive rate values on select ECG waveforms and fixed a waveform naming typo
- Added configurable sound settings (volume, alarm, heartbeat) within events, including scenario saving, defaults, and clear display states
- Improved auscultation menu behavior to prevent overlap and ensure proper closing during navigation
- Fixed layout issues with tab labels in languages with longer text
- Added collapsible folders in the Scenario Library
- Improved connection handling between Control and Monitor, including reconnection after Wi-Fi loss
- Resolved connection stability issues to prevent repeated reconnection attempts while awaiting monitor selection, eliminating device flickering
- Corrected unintended event behaviour so sound-related settings (e.g., Alarm Off) only appear when explicitly configured
LIFEPAK 35
Corrected ETCO₂ sweep speed and waveform display to better match real device behavior
Menu/Main Screen
Fixed issue where the Settings page could appear blank on some iOS devices
Revel Vent
Fixed crash when connecting Control to the monitor
SimConnect
- Updated firmware (v1.4.0) with improved logging and behaviour consistency
- Improved Wi-Fi settings retrieval during SimConnect connection, including clearer loading prompts and retrying settings requests if the router is still starting up
SimOccluder
- Updated firmware (v1.2.0) for improved performance, reliability, and stability
- Updated in-app branding
Zoll R
- Added configuration options for AED startup mode and automatic AED analysis
Zoll X
- Fixed 12-lead ECG display reverting incorrectly in Torsades rhythm
- Added CPR metronome audio feedback
- Updated ETCO₂ monitoring to include automatic calibration on activation
Multiple Screens
- Fixed missing exit prompt on monitor screens
- Prevented automatic BP from starting while the monitor is off
Maintenance and General
- Updated compatibility with the latest Apple system updates
- Standardized screen orientation (landscape by default; portrait where required)
- Fixed unit consistency in exported videos (ETCO₂ and temperature)